Table 4

Accuracy of InBody (770) and linear correction to DXA body fat percent (%BF) and fat-free mass (FFM)


Measure
InBody 770InBody correction
%BF=0.8502×InBody%BF+6.951
AllMenWomenAllMenWomen
%BF
 Bias−3.4−4.1−2.50.0−0.10.1
 MAE3.94.43.12.22.42.0
 SDE3.13.22.82.93.12.6
 LoA upper2.72.22.95.75.95.2
 LoA lower−9.5−10.3−7.9−5.7−6.0−5.0
 LoA absolute range12.212.410.911.312.010.2
FFM (kg)
 Bias−2.73.31.8−0.1−0.10.0
 MAE3.13.72.11.92.21.4
 SDE2.83.02.02.62.91.9
 LoA upper8.29.35.75.05.63.8
 LoA lower−2.8−2.6−2.2−5.1−5.8−3.8
 LoA absolute range10.911.97.910.111.47.6
  • DXA, dual-energy X-ray absorptiometry; LoA, 95% limits of agreement; MAE, mean absolute error; SDE, SD of error.